· Inflation or deflation measurement mode
· Advanced measurement algorithm
· Adoption of a wide range cuff (22-42 cm) as well as 5 other types of cuffs
· Dual-rolling pump
· Equipped with LCD display
· 5-12 V power source
· IHB screening function
· ISO:81060-2 clinical protocol validated
· Includes the well-established ECEV
· ECEV - Electric Control Exhaust Valve
A&D has newly released the TM-2917 Blood
Pressure Module, which is compatible with the previous TM-2915 model. The
TM-2917 can select either inflation or deflation measurement mode.
Features
· Inflation or deflation measurement mode
· Advanced measurement algorithm
· Adoption of a wide range cuff (22-42 cm) as
well as 5 other types of cuffs
· Dual-rolling pump
· Equipped with LCD display
· 5-12 V power source
· IHB screening function
· ISO:81060-2 clinical protocol validated
· Includes the well-established ECEV
· ECEV - Electric Control Exhaust Valve
Specifications
Item | TM-2917 | |
Blood pressure measurement | Oscillometric method | |
Operating Modes | Set either Inflation or Deflation mode | |
Patient Applications | Adult, Children, Neonatal | |
Pressure detection | For measurement: For safety circuit: | Semiconductor pressure sensor Semiconductor pressure sensor |
Measurement range | Pressure: Pulse Rate: | 0 to 299 mmHg 30 to 240 bpm |
Measurement Accuracy | Pressure: Pulse Rate: | ±3 mmHg over full range in operating conditions ±2% or ±2 BPM, whichever is greater |
Systolic Range | (Inflation) Adult: Neonatal: (Deflation) Adult: Neonatal: | 60 to 280 mmHg N/A 40 to 280 mmHg 40 to 130 mmHg |
DIA Range | (Inflation) Adult: Neonatal: (Deflation) Adult: Neonatal: | 40 to 200 mmHg N/A 20 to 200 mmHg 20 to 100 mmHg |
Scale interval | Pressure: Pulse Rate: | 1 mmHg 1 bpm |
Inflation method | Pump | Rolling pump 10 s or less/ 0→300 mmHg |
Exhaust system | Constant exhaust: Rapid exhaust: | Automatic deflation controlled by solenoid valve Rapid exhaust by opening of solenoid valve for controlled exhaust and rapid exhaust |
Electrical rating | Rated power supply voltage: Power input: | Direct current +5 to +12 V Not more than 8 VA |
Serial communication | Standardl: Logic levels: Data-flow control: | Based on RS232C TTL (3.3V) signal voltage level RTS from NIBP MODULE only (Non-use CTS) |
Operating Conditions | Temperature: Humidity: Air pressure: | 0˚C to 60˚C Not more than 85%RH (no condensation) 700 - 1060hPa |
Storage environment | Temperature: Humidity: Pressure: | -20 to +70°C 95%RH or less (with no condensation) 700 - 1060hPa |
External dimensions | 60(H) x 39(W) x 104.5(D) | |
Weight | 260g | |
NIBP Accuracy | Meets ISO 81060-2:2013 | |
Safety & Regulatory Standards | IEC 60601-1:2005+AMD1:2012, IEC 60601-1-2:2014, IEC 810601-2-30:2018 ISO 81060-2:2013, IEC 62304:2006 | |
Service life | Time Cycles | 6 years 100,000 |